Saturday, 28 December 2013

Gas Chromatography

Gas Chromatography

by Warangkana Punrattanasin and Christine Spada


Table of contents

      Introduction
        Basic Components of a GC
        Theory of Gas Chromatography
        Example Chromatograms
        EPA 500 Series Methods
        Key to Terminology
        Pop Quiz
        Web Links to Related Topics
        References

    Introduction


    Gas chromatography (GC) is an analytical technique for separating compounds based primarily on their volatilities. Gas chromatography provides both qualitative and quantitative information for individual compounds present in a sample. Compounds move through a GC column as gases, either because the compounds are normally gases or they can be heated and vaporized into a gaseous state. The compounds partition between a stationary phase, which can be either solid or liquid, and a mobile phase (gas). The differential partitioning into the stationary phase allows the compounds to be separated in time and space.



    Figure 1. Gas Chromatographic System





    Figure 2. Schematic of a Gas Chromatographic System






    Basic Components of a GC


    Gas supply or Carrier Gas


    Figure 3. Gas Supply

    The carrier gas is usually helium, hydrogen, or nitrogen. This serves as the mobile phase that moves the sample through the column. The carrier gas flow can be quantified by either linear velocity, expressed in cm/sec, or volumetric flow rate, expressed in mL/min. The linear velocity is independent of the column diameter while the flow rate is dependent on the column diameter.





    Injector


    Figure 4. Auto Sampler Injection System (Hewlett-Packard Model No. 7673)

    The injector is a hollow, heated, glass-lined cylinder where the sample is introduced into the GC. The temperature of the injector is controlled so that all components in the sample will be vaporized. The glass liner is about 4 inches long and 4 mm internal diameter.



    Column


    Figure 5. Capillary GC Column (Supelco ® PAG)

    The GC column is the heart of the system. It is coated with a stationary phase which greatly influences the separation of the compounds. The structure of the stationary phase affects the amount of time the compounds take to move through the column. Typical stationary phases are large molecular weight polysiloxane, polyethylene glycol, or polyester polymers of 0.1 to 2.5 micrometer film thickness. Columns are available in many stationary phases sizes. A typical capillary column is 15 to 60 meters in length and 0.25 to 0.32 mm ID. A typical packed column is 6 to 12 feet long and 2.2 mm ID.



    Oven


    Figure 6. GC Oven with column in place

    The column is placed in an oven where the temperature can be controlled very accurately over a wide range of temperatures. Typically, GC oven temperatures range from room temperature to 300�C, but cryogenic conditions can be used to operate at temperatures from about -20�C to 20�C.



    Detector


    Figure 7. Electron Capture Detector (ECD) and Flame Ionization Detector (FID) (Shown actual size)

    As compounds come off the column, they enter a detector. The compound and detector interact to generate a signal. The size of the signal corresponds to the amount the compound present in the sample. There are several different types of detectors that can be employed, depending on the compounds to be analyzed. These detectors can measure from 10-15 to 10-6 gram of a single component.



    Data Recorder System


    Figure 8. Data Recorder

    The data recorder plots the signal from the detector over time. This plot is called a chromatogram. The retention time, which is when the component elutes from the GC system, is qualitatively indicative of the type of compound. The data recorder also has an integrator component to calculate the area under the peaks or the height of the peak. The area or height is indicative of the amount of each component.





    Theory of Gas Chromatography


    Retention Time (tR)

    The retention time is the total time that a compound spends in both the mobile phase and stationary phase. Retention time is generally reported in minutes.

    Dead Time (tm)

    The dead time is the time a non-retained compound spends in the mobile phase which is also the amount of time the non-retained compound spends in the column. Dead time is generally reported in minutes.

    Adjusted Retention Time (tR')

    The adjusted retention time is the time a compound spends in the stationary phase. The adjusted retention time is the difference between the dead time and the retention time for a compound.


    Capacity Factor (or Partition Ratio) (k')

    The capacity factor is the ratio of the mass of the compound in the stationary phase relative to the mass of the compound in the mobile phase. The capacity factor is a unitless measure of the column's retention of a compound.


    Phase Ratio (ß)

    The phase ratio relates the column diameter and film thickness of the stationary phase. The phase ratio is unitless and constant for a particular column and represent the volume ratioß.


    Distribution Constant (KD)

    The distribution constant is a ratio of the concentration of a compound in the stationary phase relative to the concentration of the compound in the mobile phase. The distribution constant is constant for a certain compound, stationary phase, and column temperature.


    Selectivity (or Separation Factor) (alpha)

    The selectivity is a ratio of the capacity factors of two peaks. The selectivity is always equal to or greater than one. If the selectivity equals one the two compounds cannot be separated. The higher the selectivity, the more separation between two compounds or peaks.


    Linear Velocity (u)

    The linear velocity is the speed at which the carrier gas or mobile phase travels through the column. The linear velocity is generally expressed in centimeters per second.


    Efficiency

    The efficiency is related to the number of compounds that can separated by the column. The efficiency is expressed as the number of theoretical plates (N, unitless) or as the height equivalent to a theoretical plate (HETP, generally in millimeters). The efficiency increases as the height equivalent to a theoretical plate decreases, thus more compounds can be separated by the column. The efficiency increases as the number of theoretical plates increases, thus the column's ability to separate two closely eluting peaks increases.




    Example Chromatograms


    Sample contains 6 aromatic hydrocarbons dissolved in a solvent (methanol). The compounds' properties are summarized in Table 2.


    The compounds were separated on an nonpolar, 95% methyl, 5% phenylpolysiloxane column, 30 m long, 0.25 mm ID, and 0.25 micrometer film thickness. About 1 microliter of the hydrocarbon sample were injected. Approximately 5 nanograms (ng) of each component was injected per 1 microliter. A flame ionization detector (FID) was used.

    Temperature Programming Effects


    Figure 9.

  1. Temperature Program: 50�C (min) - 10�C/min - 100�C
  2. Head Pressure: 12 psi
  3. Split Ratio: 1/50
  4. This chromatogram shows an ideal temperature program for separation of the 6 aromatic compounds on this column. The first peak is the solvent, methanol. The compounds elute in order of increasing boiling point, that is, compounds with higher boiling points are more retained by the stationary phase. Note that para-xylene and meta-xylene cannot be separated on this column; the peak (#5) containing these compounds is broad at the baseline and shows a distinct shoulder.




    Figure 10.

  5. Temperature Program: 60�C Isothermal
  6. Head Pressure: 12 psi
  7. Split Ratio: 1/50
  8. This chromatogram shows the effects of an isothermal* temperature program at 60�C. The result is an increase in the retention time of all compounds. The heights of the later eluting peaks are reduced and the peak widths increased because they are more affected by the lower temperature program used. (*isothermal means a constant oven temperature was used throughout the run.)





    Flow Rate Effects


    Figure 11.

  9. Temperature Program: 50�C (1 min) - 10�C/min - 100�C
  10. Head Pressure: 9 psi
  11. Split Ratio: 1/50
  12. This chromatogram shows the effects of a reduced head pressure while using the ideal temperature program. The flow rate was reduced by decreasing the head pressure. The retention time is slightly increased due to the low flow rate used. All of the peak heights were reduced and the peak widths are increased.



    Figure 12.

  13. Temperature Program: 50�C (1 min) - 10�C/min - 100�C
  14. Head Pressure: 15 psi
  15. Split Ratio: 1/50
  16. This chromatogram shows the effects of a higher head pressure while using the ideal temperature program. The flow rate was increased by increasing the head pressure. The retention time was reduced and all of the peak heights were increased.




    Split Ratio Effects


    Figure 13.

  17. Temperature Program: 50�C (1 min) - 10�C/min - 100�C
  18. Head Pressure: 12 psi
  19. Split Ratio: 1/25
  20. This chromatogram shows the effects of a low split ratio while using the ideal temperature program. All of the peak heights were increased due to the greater amount of the sample introduced into the column.




    Figure 14.

  21. Temperature Program: 50�C (1 min) - 10�C/min - 100�C
  22. Head Pressure: 12 psi
  23. Split Ratio: 1/75
  24. This chromatogram shows the effects of a high split ratio while using the ideal temperature program. All of the peak heights were reduced due to the smaller amount of the sample introduced into the column.





    Effect of Stationary Phase on Separation of Para-Xylene and Meta-Xylene


    Figure 15.

    This chromatogram shows the separation of benzene, toluene, para-xylene, meta-xylene and ortho-xylene. The first peak is the solvent, hexane. A polyalkylene glycol fused silica capillary column 30 m long, 0.25 mm ID, and 0.25 micrometer film thickness was used for separation. Para-xylene (peak #4) and meta-xylene (peak #5) can be separated on this column. This illustrates the matching of the stationary phase with the desired compounds to be separated.







    EPA 500 Series Methods


    • Method 502.1: Volatile haloginated organic compounds in water by purge and trap gas chromatography
    • Method 502.2: Volatile organic compounds in water by purge and trap capillary column gas chromatography with photoionization and electrolytic conductivity detectors in series
    • Method 503.1: Volatile aromatic and unsaturated organic compounds in water by purge and trap gas chromatography
    • Method 504: 1,2-dibromoethane (EDB) and 1,2-dibromo-3-chloropropane (DBCP) in water by microextraction and gas chromatography
    • Method 505: Analysis of organohaline pesticides and aroclors in drinking water by microextraction and gas chromatography
    • Method 507: Determination of nitrogen-and phosphorus-containing pesticides in water by gas chromatography with a nitrogen-phosphorus detector
    • Method 508: Determination of chlorinated pesticides in water by gas chromatography with an electron capture detector
    • Method 510.1: Determination of the maximum total trihalomethane potential
    • Method 515: Determination of chlorinated herbicides in drinking water
    • Method 524.1: Volatile organic compounds in water by purge and trap gas chromatography/ mass spectrometry
    • Method 524.2: Volatile organic compounds in water by purge and trap capillary column gas chromatography/ mass spectrometry
    • Method 525: Determination of organic compounds in drinking water by liquid-solid extraction and capillary column gas chromatography/ mass spectrometry



    Key to Terminology








    Pop Quiz



    1. What is the component that most influences the separation of compounds?
    a) injector
    b) mobile phase
    c) stationary phase

    2. Given napthalene (boiling point = 218�C), phenol (boiling point = 181.7�C), and toluene (boiling point = 110.6�C). Which compound will elute first on a nonpolar column?
    a) naphthalene
    b) phenol
    c) toluene

    3. What will happen to the retention time if the flow rate is increased?
    a) increase
    b) decrease
    c) no change

    4. If a variable temperature program is used rather than an isothermal temperature program, what parameter will not be affected?
    a) order in which compounds elute
    b) retention time
    c) peak height

    5. What will happen to the peak height if the split ratio is decreased?
    a) increase
    b) decrease
    c) no change

    1 comment: