Sunday 10 November 2019

Selecting a suitable detector for Gas Chromatographic analysis

Selecting a suitable detector for Gas Chromatographic analysis

Gas Chromatography offers a wide range of detectors to meet your analysis applications. The two detectors which are in common use are the Flame Ionisation detector and Thermal Conductivity detector as these cover detection of majority of compounds encountered in routine applications. However, there are other detectors also available which are useful for specific applications.
Schematic of Flame Ionization Detector
Schematic of Flame Ionization Detector
Gas Chromatographic detectors fall under three distinct categories:
Non –selective-Respond to all compounds coming out of the chromatographic column except the carrier gas.
Selective– Respond to a range of compounds with common physical or chemical characteristics
Specific-respond only to a single chemical compound in the carrier stream
Desirable features of a detector
The desirable features in the detector selected for your application should cover some if not all of the following attributes:
  • Reproducibility of response to changes in composition of eluting gas
  • High sensitivity
  • Large linearity of response
  • Least peak broadening and resultant loss of resolution
  • Preferably non-destructive
Detector types and their common applications
Flame Ionisation Detector (FID)
FID is the most popular of all detectors. It shows high sensitivity to most organic compounds. The advantages of FID are high sensitivity, large linear response, low noise and ruggedness. However, it is destructive in nature and the sample gets destroyed. Compounds with little or no response include inorganic gases such as CO, CO2, NH3, CS2, NOX, noble gases, halogenated compounds, formic acid and formaldehyde.
Thermal Conductivity detector (TCD)
TCD is also referred to as a universal detector as it responds to most compounds including inorganic gases such as O2, N2 and CO2.It permits detection at both high percent and low ppm levels.
Nitrogen Phosphorus Detector (NPD)
The Nitrogen Phosphorus detector responds selectively to most organic compounds that contain phosphorus or organically bound nitrogen down to the picogram level. Its high level of sensitivity and specificity makes it suitable for analysis of low levels of drugs and pesticides.
Electron Capture Detector (ECD)
Electron Capture detector provides trace analysis of halogenated compounds, nitriles, nitrates and compounds with conjugated double bonds. It finds widespread use in analysis of environmental samples containing organochlorine pesticides, PCB’s etc.
Flame Photometric Detector (FPD)
Flame photometric detector is a highly sensitive detector for phosphorus   compounds or alkyl tin compounds. Sulfur compounds include mercaptans, alkyl sulphides, H2S, CS2, SO2 andCO2 in petroleum fractions and pulp-milling processes. Alkyl tin finds use in marine antifouling paints which are applied to boat hulls and ocean oil rigs. Such paints released into the sea pose an environment hazard.
Photoionization detector (PID)
The photoionization detector responds to compounds ionized by UV light .it is used to detect aromatic and unsaturated compounds in drinking water, waste water, soil and sludge samples. Other applications include detection of aromatics such as benzene, toluene, xylene and polynuclear aromatic hydrocarbons in petroleum fractions.
Detector Selection Summary
DetectorTypical Classes of CompoundsDetection limitsDynamic range
FIDMajority of organic compounds100 pg107
TCDUniversal detector1ng107
ECDHalides, nitrates, nitriles, peroxides, anhydrides, organometallics50fg105
NPDN, P containing compounds10pg106
FPDS,P containing compounds100pg103
PIDAromatics, Polynuclear aromatic hydrocarbons2pg107

No comments:

Post a Comment