Contamination From Reagents
Trace Analysis Guide: Part 9 By Paul Gaines, Ph.D.
High Purity Water
Sample preparations using acid digestion, fusion, or ashing all typically use water as the primary reagent. Most water used in trace metal laboratories is produced by systems that use ion-exchange purification. This water is commonly referred to as "conductivity water" because its conductivity approaches the theoretical conductivity of water (0.055 microhm / cm {18.2 megohm water} at 25°C).Under normal laboratory conditions, conductivity water never measures to be 18.2 megohm due to the presence of CO2 (H2CO3HCO3- + H+). Furthermore, if you could find water giving a conductivity of 18.2 megohm, it is not necessarily free of trace elemental contaminants because only ionized compounds are detected by conductivity measurement.
Through carefully controlled experiments and measurements in clean room facilities, we have found that conductivity water will typically give readings closer to 16 megohm and that it is free of trace metallic impurities down to conventional ICP-MS and axial view ICP-OES detection limits. We have also found that sub-ppb level impurities that were once thought to be coming from the water, are in actuality from the atmosphere and the container materials (see earlier parts of this guide). We have found that the use of clean room facilities and high temperature nitric acid-leached LDPE bottles are necessary for reliably measuring common contaminant elements in water. Therefore, do not assume that your water has significant levels of elemental contaminants if it gives conductivity readings between 16 and 17 megohm and your ICP-MS or OES is detecting trace levels of the common environmental contaminants.
Storage of High Purity Water
High purity water should be used ASAP. "Stored" high purity water may pick up impurities from the storage container. Popular storage containers are made from quartz, polyethylene (both high and low density), and fluoropolymers.Quartz
Quartz (fused quartz or vitreous silica) typically contains 98.8% SiO2, and impurities consisting mainly of Na2O, Al2O3, Fe2O3, MgO, and TiO2. Quartz has a solubility in water of 11 ppm.1 We have measured a solubility of quartz in conductivity water of 11.2 ppm as silicic acid (equilibration time is ~4 weeks using 400 mesh quartz powder). More about quartzHDPE and LDPE
A significant amount of HDPE is manufactured using alumina / silica based catalysts. Long term storage in high density polyethylene (HDPE) can result in ppm levels of Ca, Mg, Si, Ti, Al and ppb levels of Cr, V and Fe. LDPE can be manufactured using an organic catalyst. Storage in HNO3 leached LDPE is optimum. Through study, we've discovered that short term (1-5 days) storage in both 20 liter HDPE and LDPE cubi containers that have been leached with dilute HNO3 do not leach any elements at ICP-MS / OES detection limits.Fluoropolymers
Fluoropolymers are not as clean as generally thought (see below figures). Studies performed in our own laboratories confirm these results. It is our recommendation that you save your money and use LDPE.Figures 9.1 and 9.2 show total trace metals and major contributing ions for cut parts and from PFA 1 resin pellets and extruded tubing following extraction.2
- Extracted for 5 days in 10% ultrapure HNO3 at 25°C -
Fig 9.2: Total Trace Metals & Major Contributing Ions from PFA 1 Resin Pellets and Tubing
Extracted for 5 days in 2% ultrapure HNO3 at 25°C
Extracted for 5 days in 18 MΩ Di H2O at 85°C for 1 hour
High Purity Acids
High purity acids have been commercially available for years with the major impurities typically less than 1 ppb. Distilling the acid yourself may offer some improvement in purity. If you are using more than 500 mL of mineral acid per month, you may want to consider the monetary savings of distilling the acid yourself. Diagrams 9.1 and 9.2 below show typical quartz and teflon stills.Diagram 9.2: All-Teflon Sub-Boiling Still
Sub-boling distilled (ng/g) | ACS Reagent-grade acid (ng/g) | |
Pb | 0.07 | 0.5 |
Tl | 0.01 | 0.1 |
Ba | 0.04 | 2 |
Te | 0.01 | 0.1 |
Sn | 0.05 | 0.07 |
In | 0.01 | - |
Cd | 0.02 | 0.03 |
Ag | 0.03 | 0.05 |
Sr | 0.01 | 0.05 |
Zn | 0.2 | 2 |
Cu | 0.1 | 4 |
Ni | 0.2 | 6 |
Fe | 3 | 20 |
Cr | 0.3 | 2 |
Sub-boling distilled (ng/g) | ACS Reagent-grade acid (ng/g) | Commercial high purity (ng/g) | |
Pb | 0.02 | 0.2 | 0.3 |
Tl | - | 0.2 | - |
Ba | 0.01 | 8 | - |
Te | 0.01 | 0.1 | - |
Sn | 0.01 | 0.1 | 1 |
In | 0.01 | - | - |
Cd | 0.01 | 0.1 | 0.2 |
Ag | 0.01 | 0.03 | 0.1 |
Sr | 0.01 | 2 | - |
Zn | 0.04 | 4 | 8 |
Cu | 0.04 | 20 | 4 |
Ni | 0.05 | 20 | 3 |
Fe | 0.3 | 24 | 55 |
Cr | 0.05 | 6 | 130 |
Ca | 0.2 | 30 | 30 |
K | 0.2 | 10 | 11 |
Mg | 0.1 | 13 | - |
Na | 1 | 80 | - |
Total Impurity | 2.3 ppb | 220 ppb | 220 ppb |
Sub-boling distilled (ng/g) | ACS Reagent-grade acid (ng/g) | Commercial high purity (ng/g) | |
Pb | 0.2 | 2 | 16 |
Tl | 0.1 | 0.1 | - |
Ba | 0.1 | >1000 | 10 |
Te | 0.05 | 0.05 | - |
Sn | 0.3 | 0.03 | <1 |
Cd | 5 | 0.1 | 4 |
Ag | 1 | 0.1 | 0.5 |
Sr | 0.02 | 14 | - |
Zn | 0.1 | 7 | 17 |
Cu | 0.1 | 11 | 3 |
Ni | 0.5 | 8 | 0.5 |
Fe | 2 | 330 | 10 |
Cr | 9 | 10 | 18 |
Ca | 0.2 | 760 | 7 |
K | 0.6 | 200 | 9 |
Mg | 0.2 | 500 | 4 |
Na | 2 | 600 | - |
Total Impurity | 16 ppb | >3400 ppb | 100 ppb |
Sub-boling distilled (ng/g) | ACS Reagent-grade acid (ng/g) | |
Pb | 0.6 | 0.5 |
Tl | 0.1 | 0.1 |
Ba | 0.3 | 0.2 |
Te | 0.1 | 0.1 |
Sn | 0.2 | 0.6 |
Cd | 0.3 | 0.2 |
Ag | 0.3 | 0.6 |
Sr | 0.3 | 0.4 |
Zn | 0.5 | 2 |
Cu | 0.2 | 6 |
Ni | 0.2 | 0.5 |
Fe | 7 | 6 |
Cr | 0.2 | 0.2 |
Ca | 2 | 123 |
K | 4 | 9 |
Mg | 2 | 4 |
Na | 9 | 50 |
Total Impurity | 27 ppb | 200 ppb |
Sub-boling distilled (ng/g) | ACS Reagent-grade acid (ng/g) | |
Pb | 0.05 | 0.8 |
Tl | 0.1 | 0.2 |
Ba | 0.1 | 0.5 |
Te | 0.05 | 0.1 |
Sn | 0.05 | 11 |
Cd | 0.03 | 2 |
Ag | 0.05 | 0.1 |
Sr | 0.1 | 0.5 |
Zn | 0.2 | 4 |
Cu | 0.2 | 3 |
Ni | 0.3 | 12 |
Fe | 0.6 | 110 |
Cr | 5 | 20 |
Ca | 5 | 14 |
K | 1 | 28 |
Mg | 2 | 10 |
Na | 2 | 100 |
Total Impurity | 17 ppb | 320 ppb |
Other Reagents
The "purity" situation for salts and other reagents is typically not as favorable as it is for water and acids. Typically, the highest purity solid that can be confirmed is 99.999% (5-9's or TMI 10ppm). This translates to 100 ppb total impurities for a 100 fold dilution (i.e. - 1 g of 5-9's pure reagent into 100 mL of solution). We recommend that you know your supplier's definition of "high purity".
CAUTION: Do NOT use solid reagent grade materials when preparing samples for trace metals analysis whenever possible.
If
necessary, a blank should be performed to confirm the acceptability of
reagents(s) or to identify problem elements / impurities. Purification
of unacceptable reagents may be accomplished by mercury cathode
electrolysis, extraction with dithiozone or cupferron, ion-exchange, or crystallization.
1. Ralph K. Iler, The Chemistry of Silica, John Wiley & Sons: New York, (1979).