Thursday 24 December 2015

What is laminar air flow system?



What is laminar air flow system?
A laminar flow cabinet or laminar flow closet or tissue culture hood is a carefully enclosed bench designed to prevent contamination of semiconductor wafers, biological samples, or any particle sensitive materials. Air is drawn through a HEPA filter and blown in a very smooth, laminar flow towards the user.
What is a biological safety cabinet?
A biosafety cabinet (BSC) — also called a biological safety cabinet or microbiological safety cabinet — is an enclosed, ventilated laboratory workspace for safely working with materials contaminated with (or potentially contaminated with) pathogens requiring a defined biosafety level.
What is the biosafety?
Biosafety is the prevention of large-scale loss of biological integrity, focusing both on ecology and human health. These prevention mechanisms include conduction of regular reviews of the biosafety in laboratory settings, as well as strict guidelines to follow. Biosafety is used to protect us from harmful incidents.
Thursday, Dec 24, 2015 Get the latest updates from us for free

SOP for Laminar Air Flow Unit

1.0  OBJECTIVE:
       To lay down a Procedure is to provide guidelines for operation of Laminar Air Flow.

2.0  SCOPE:
       This procedure is applicable for LAF which is installed in Microbiology laboratory of QC Department
3.0  RESPONSIBILITY:
       Microbiologist
4.0  ACCOUNTABILITY:
       HEAD QA & QC
5.0  PROCEDURE:
5.1  Operation & Cleaning of LAF
5.1.1  Switch “OFF” the UV light.
5.1.2  Clean the equipment with clean lint free cloth duster and spray 70% IPA solution to LAF unit and entire Area of LAF room.
5.1.3  Switch “ON” the “AIR” and Visible Light switch situated on the control panel on right hand side of the instrument.
5.1.4  Now check the LAF Manometer pressure, it should be within 10 – 20 mm of Water Gauge.
5.1.5  Check the log record for burning hour of UV light, it should not exceed more than 2000 hours. UV tube should be replaced after specified burning period.
5.1.6  Start the gas burner with the help of gas lighter and carry out the routine works.
5.1.7  After completion of work, switch OFF the Airflow and gas burner.
5.1.8  Clean any remaining water or waste liquid material spilled on the laminar airflow platform properly with a dry cloth.
5.1.9  Clean properly the working chamber from ceiling, followed by side glass and then the platform of the chamber with sterile 70% IPA.
5.1.10  Spray sterile 70% IPA and switch OFF the Visible light and then Switch ON the UV light till to start the next operation.
5.1.11  If the LAF is used for Aseptic filtration, clean the receiver pipe tank with hot WFI.
5.1.12  Open the outlet of Solution Collection Tank and collect the solution in plastic crate. Close the valve and transfer the crate for solution disposal.
5.2  Cleaning of Pre filters
5.2.1  Ensure that the LAF is switch off. Affix the tag on Instrument “UNDER MAINTENANCE”.
5.2.2  Pre filter is situated on the backside of the LAF, unscrew and then remove the pre filter from LAF.
5.2.3  Transfer the pre filter to washing area and blow the compressed air from reverse side to blow out all the dust particles.
5.2.4  Wash the pre filter with DM water and then with liquid detergent solution. Finally wash again with DM water to remove the detergent solution.
5.2.5  Remove the trapped water from the pre filter by jerking and finally rinse with 70% IPA solution.
5.2.6  Allow to dry the pre filter at room temperature.
5.2.7  Assemble the pre-filter at their original place and tighten the screw properly.
5.3  Precautions
5.3.1  Take care to prevent any damage to the integrity of filter during cleaning. In case of observation of any damage to filters, immediately inform to your superior for further action.
5.3.2  Instrument should be cleaned when the electrical connections of the equipment is in OFF position.
5.3.3  Clean the laminar airflow chamber after every operation.
5.3.4  Do not work when UV light is ON as it may cause eye damage.
5.3.5  Maintain the level of the platform by adjusting the equipment from the base with the help of glass beads in such a manner that the platform level should remain horizontally flat. There should be no ups and downs in any portion of the chamber.
5.4  Frequencies
5.4.1  Daily for operation
5.4.2  Fortnightly for Pre filter cleaning
5.4.3  Calibration: Yearly by External agency
6.0  ABBREVIATIONS:
SOP : Standard Operating Procedure
QA : Quality Assurance
QC : Quality Control
NA : Not Applicable
LAF : Laminar Air Flow
IPA : Isopropyl Alcohol
DM : Dematerialized

3 comments:

  1. I am very grateful to read your blog. Thank for the share

    Acmas Technologies Inc is leading laminar airflow cabinet or vertical autoclave manufacturer in India.



    ReplyDelete
  2. Hello, really liked your page. Thanks for the post. I Really Found this topic very interesting and useful.

    If you want to know more about the Laminar Air Flow, please visit our website:

    ReplyDelete
  3. Starscientific is the leading fume hood manufacturers in India. We provide kinds of Fume Hoods for the Laboratory. The provided fume hood is manufactured using high-grade components with the help of contemporary technology under the surveillance of adroit professionals at our well-equipped production house. Laboratory fume hoods are one of the most reliable engineering controls in the laboratory. Fume hoods protect lab users by containing vapors, dust, and gases generated within the hood. Fume hoods remove the fumes as air flows into the hood and then out via the laboratory exhaust system.
    you may also visit:

    laminar air flow manufacturers,
    stability manufacturers,
    deep freezer - 80'c manufacturers
    Mortuary Chamber Manufacturers
    tray dryer manufacturers in Delhi
    tray dryer manufacturers in Delhi
    press box manufacturers

    Humidity chamber manufacturers
    Biosafety manufacturers

    ReplyDelete