Tuesday 24 July 2012

Analytical Balance Calibration


Analytical Balance Calibration

Calibration procedure:

Internal calibration

It shall be performed as per the manufacturer’s instructions.
Drift check
For the calculation of the drift 10 mg weight shall be used.
Note down the 10 measurements of the 10 mg weight in the performance check log. ( Attachment no. 01)
Variation in the observed weight from the mean value does not exceeds  ± 0.2 mg.
The 10 mg weight shall meet the performance check criteria of the mass value i.e. 0.1% of actual mass value.
E.g. For all the 10 measurements of 10 mg weight, variation in the weighings cannot exceed 0.01mg.
Performnce check
After the Auto calibration put 1 mg, 2 mg, 5 mg, 10 mg and 20 mg weights individually. The measurement shall be within the 0.1% of actual mass value of the individual weight as given in the performance check log.( Attachment no. 02)
Measurement uncertainty check
The measurement of uncertainty shall be carried out by using 10 mg weight.
Put the external weight of 10 mg on the pan and note the 10 measurements. (Attachment no. 03)
Calculate the measurement of uncertainty as follows.
Measurement of uncertainty shall be Not More Than 0.001.
CALIBRATION OF TOP PAN BALANCE
Calibration frequency: It shall be performed daily, after any maintenance, after the relocation and after the power failure of the balance.Each manufacturer supplies the maximum loading condition of the balance. The calibration of the balance shall be designed in such a way that the performance check weights cover the entire loading range of the balance. Lower and higher load limits shall be checked for the performance check. Before calibration make sure that the level bubble is in the center of the indicator and if internal calibration facility is available in the balance then it shall also be performed daily before performance check of the balance.


Always use a check weight appropriate for the particular balance. E.g. a 100 – mg check weight might be selected for a balance that has a load limit of 150 mg. Thus the operator must know the maximum capacity of the balance to select the correct check weight.

Attachment-1: Performance Check Log of Microbalance
Instrument:_____________________                                                        Code No :______________

Make         :_____________________                                                       Model   :_______________ 

Date
Level indicator check
1 mg

2 mg
5 mg
10 mg
20 mg
Weight box no. and valid upto
Done by
Checked
by
Remark
Std. Wt. Cal No. Þ






Limit
Þ
Bubble should be in Center
+ 0.1% of
 Std. Weight
(        to        )
+ 0.1% of
 Std. Weight
(        to        )
+ 0.1% of
Std. Weight
(        to        )
+ 0.1% of Std. Weight
(        to        )
+ 0.1% of Std. Weight
(        to        )
























Attachment-2: Auto Performance Check Log of Microbalance
Instrument:_____________________                                                        Code No :______________
Make         :_____________________                                                       Model   :_______________ 
Check level indicator (Bubble should be in center) : _________________________________









                                                               (Paste Performance check chart here)
       
Calibrated by : _______________                                               Date : ____________________
Checked  by : _______________                                                 Date : ____________________
         
Attachment-3: Measurement Uncertainity Check of Microbalance

Instrument:_____________________                                                         Code No :______________
Make         :_____________________                                                        Model   :_______________ 

Date
Std. Wt.
Std. Wt. Cal. No.
Observation
Observation of Uncertainty
Less than or equal to 0.001
Weight box no. and valid upto
Done by
Checked  by
Remark

10 mg

















Calculation

No comments:

Post a Comment