Good Laboratory Practices
• Methods
Ø Use
/ follow correct specifications, methods of analysis and SOPs at the time of work.
• Personnel
Ø Only
trained persons should perform the activity.
• Instruments/Equipments
Ø Qualified & calibrated instruments / Equipments to be used.
Ø While working in laboratory, ensure that all the required PPE’s & RPE’s are used.
Ø Ensure that the instruments are in calibrated status.
Ø After completion of analysis if the instrument is not in use, then ensure that all
switches are turned off.
Ø All the distillation and reflux operations should be performed in hood and the
analyst should attend to the experiment.
Ø Clean the working table, instruments before & after completion of analysis.
Ø Keep
the laboratory notebook, column and chemicals at designated places after / when
not in use.
Ø After opening chemical bottle, write down the opened on date and use before date on
the label of the bottle.
Ø Take all necessary precautions while handling Hazardous/poisonous chemicals.
Ø Follow proper garment policy while entering in the microbiology laboratory.
Ø Do not enter in the areas which are restricted for entrance.
Ø Do not bring any food/ medicines in the laboratory.
• Dilutions:
Ø Use appropriate, cleaned & dried volumetric apparatus.
Ø Label the glassware, which is in use.
Ø Ensure that the pipettes and volumetric flasks are of Class A grade.
Ø Ensure that the tip of pipettes is proper and not broken.
Ø Use appropriate solvents for dilution.
Ø Wipe the pipette before inserting in the volumetric flask.
Ø Make the dilutions carefully & accurately.
Ø Do not use contaminated sucker / rubber bulb.
Ø Make up the volume of the solution when it is at room temperature.
Ø Follow the methods properly & crosscheck once.
• Weighing:
Ø Ensure that the balances are clean and calibrated.
Ø Check the spirit level of the balance before use.
Ø Balance weighing range & capacity should be considered before weighing.
Ø Take the weight once the balance is stabilized.
Ø During weighing the door of the balance & balance room should be closed properly
Ø Follow the SOP correctly.
Ø Avoid spillage.
Ø Do not forget to check cleanliness before & after weighing.
• Mobile phase preparation:
Ø Use appropriate solutes & solvents as per the specified procedure.
Ø The solvents should be degassed before use.
Ø Weigh as accurate as possible.
Ø Make pH adjustments on calibrated pH meter & always within the tolerance given
in the method.
Ø The pH meter should be calibrated for upper and lower buffer pH range.
Ø Do not forget to clean the pH meter electrode with plenty of water before and
after use.
Ø Make appropriate entries of the pH calibration and buffer in the pH meter log.
Ø Take appropriate measuring cylinder for mobile phase preparation.
Ø Filter & degas before use.
Ø Prepare sufficient mobile phase required for analysis. Do not mix two lots of mobile
phase.
• Column care:
Ø Avoid any jerks or vibrations to the column.
Ø Always replace the column ferrules whenever column is to be kept idle & fix end
caps to avoid silica drying.
Ø Store the column in appropriate solvents suggested by manufacturer.
Ø Wherever specified, use proper guard columns.
Ø Use the column for particular Raw material/ product/ test whenever required.
Ø Before and after analysis wash the column with prescribed solvent.
Ø Use/ Attach the column in same direction of flow provided by manufacturer.
·
HPLC
Ø Flush
the complete system (without column) with warm water (40°) once in a week using
union instead of column.
Ø If
the system suitability is not achieved by the column then regenerate the column
as per manufacturer’s instructions / SOP for column generation.
Ø Purge
seal wash line, injector and all the inlets of pump with water before use.
Ø Sonicate
all ports inlet filter with [water: methanol (50:50)] once in a week.
Ø Check
the tubing/Syringe for air bubbles of all ports.
Ø Check
the valves for precipitation & clogging.
Ø Clean
all the inline filters, frits & check valve cartridges at least twice in a
month.
Ø Before
converting the system to normal phase wash with water/ Methanol/ IPA/ Diluent
serially.
Ø After
completion of analysis, follow reverse procedure to convert to reverse phase.
Ø Do not store system in solvents.
Ø For purging of system start with lower flow
and then stabilize system at required condition.
Ø After analysis take backup of the data in the relevant
location.
Ø Ensure
that the system is calibrated before use.
Ø Before
start, ensure that respective solvents & inlets inserted are as per the
requirements.
Ø After
analysis, do not leave the system idle in harsh solvents like ACN, store the
system in ACN : Water mixture (20:80).
Ø Avoid
spillage on the system.
Ø Do
not interchange any of the modules or parts amongst the systems even of the
same make.
No comments:
Post a Comment