Friday 28 December 2012

Total Phosphorus




TABLE OF CONTENTS

Total Phosphorus

                                                                                                                                                     Page

Section 1:  Introduction to Total Phosphorus................................................................................        2
Section 2:  Glossary....................................................................................................................        2
Section 3:  Safety and Hygiene....................................................................................................        2
Section 4:  Sampling...................................................................................................................        3
Section 5:  Sample Preservation..................................................................................................        3
Section 6:  Sample Containers.....................................................................................................        3
Section 7:  Interpretations............................................................................................................        3
Quiz 8.1......................................................................................................................................        4
Section 8:  Total Phosphorus.......................................................................................................        4
Section 9:  Equipment and Reagents............................................................................................        4
Section 10:  Laboratory Procedure...............................................................................................      4-5
Section 11:  Total Phosphorus.....................................................................................................        5
Section 12:  Equipment and Reagents..........................................................................................        5
Section 13:  Laboratory Procedure...............................................................................................      5-6
Section 14:  Calibration Curve......................................................................................................        6
Section 15:  Calculations.............................................................................................................      6-7
Section 16:  Interferences............................................................................................................        7
Section 17:  QA/QC.....................................................................................................................      7-8
Quiz 8.2......................................................................................................................................        8
Answers to Quizzes.....................................................................................................................      8-9
Appendix A:  References
Appendix B:  Phosphorus Reagents
Appendix C:  Sample Bench Sheet
Appendix D:  Preparation of a Calibration Curve
Appendix E:  Methods Checklist
Appendix F:  Method Checklist - Reagent
Appendix G:  Method Checklist Digester
Appendix H:  Phosphorus Ascorbic Acid (Single Reagent)



TOTAL PHOSPHORUS


Section 1:  INTRODUCTION TO TOTAL PHOSPHORUS


Wastewater is relatively rich in phosphorus compounds.  Phosphorus is a nutrient used by organisms for growth.  It occurs in natural water and wastewater bound to oxygen to form phosphates.  Phosphates come from a variety of sources including agricultural fertilizers, domestic wastewater, detergents, industrial process wastes and geological formations.

The discharge of wastewater containing phosphorus may cause algae growth in quantities sufficient to cause taste and odor problems in drinking water supplies.  Dead and decaying algae can cause oxygen depletion problems which in turn can kill fish and other aquatic organisms in streams.  For this reason, phosphorus removal is an essential role of wastewater treatment plants and testing for phosphorus in the plant effluent is critical.

Phosphates are classified as orthophosphates, polyphosphates and organic phosphates.  In this procedure, orthophosphates can be determined directly by colorimetric analysis.  Other types require a digestion step to convert the “combined” phosphate to the ortho form for analysis.  This gives the Total Phosphorus result.

Section 2:  GLOSSARY


Blank:  A preliminary analysis omitting only the sample to provide an unbiased reference point or baseline for comparison.  The Blank is usually run on distilled water.

Nutrient:  Any substance used by living things that promotes growth.  The term is generally applied to nitrogen and phosphorus in water, but is also applied to other essential and trace elements.

Phosphate curve:  A calibration curve that plots concentrations of known standards against the instrument’s response (color intensity, absorbance, millivolts, etc.).

Section 3:  SAFETY AND HYGIENE

Whenever samples of wastewater are handled, it is very important that operators wash his or her hands before eating or smoking.  While some laboratory chemicals are not dangerous, many of them are poisonous or harmful to skin and clothing.  Rubber gloves and safety glasses should be used.  It is important to wash thoroughly with soap and water after handling laboratory chemicals, especially if chemicals come into contact with the skin.  Keep bench areas free of clutter and clean bench surfaces with disinfectant after testing.

Read the labels carefully and know what to do in case of an accidental spill.  Always clean up spills quickly and in the safest possible manner using disposable rags or towels.

Acids and bases can be corrosive.  Care should be taken when handling them.  Never add water to acid.  Always add acid to water very slowly because acids and bases can generate heat when mixed with water.

The Total Phosphorus test involves heating samples in the digestion step.  To avoid injuries caused by heat, use gloves or tongs to handle hot or potentially hot glassware.  Never assume that a beaker or flask is cool.

Section 4:  SAMPLING


Samples used for the determination of phosphorus can be either grab or composite.  The type of sample used will depend on the monitoring requirements, plant operating procedures and the testing and sample storage capabilities of individual plants.  Samples should be collected from well-mixed areas in the process flow so that they are representative of the total flow.

Section 5:  SAMPLE PRESERVATION


If Total Phosphorus is the only determination to be made, acidify the sample to less than pH 2 with sulfuric acid and cool to 4°C.  Acidified and cooled samples may be held for up to 28 days.  If the test is run immediately, no preservative is required.

Samples to be tested for dissolved reactive phosphorus, dissolved acid-hydrolyzable phosphorus, or total dissolved phosphorus must be filtered.  See the 18thth Edition of “Standard Methods for the Examination of Water and Wastewater” for information on these tests.

Section 6: SAMPLE CONTAINERS

Special sampling devices and storage containers are not necessary for phosphorus testing.  Sampling devices should draw from well-mixed areas of tanks or pipes, be made of resistant materials that will not rust or corrode, be capable of taking samples that are proportional to the plant’s flow, and be easily and thoroughly cleaned.

Storage containers should be made of corrosion resistant material which can stand repeated refrigeration.  These containers should have leak-proof tops.  Do not store low concentration samples in plastic bottles since phosphorus may be adsorbed on the container walls.

Rinse all sampling containers with diluted (1 + 1) hydrochloric acid.  Then rinse 3 times with distilled water.

Never use commercial detergents for cleansing any glassware used in the storage or analysis of samples for phosphorus determination.

 

Section 7:  INTERPRETATIONS


Because of its effect on receiving streams, testing for phosphorus is critical to wastewater treatment plants.  Ranges for phosphorus concentrations are given below.

               Sample type                                      range
               Influent                                              5    -   10     mg/L
               Primary Effluent                                 4    -     9     mg/L
               Activated Sludge Effluent                   3    -     8     mg/L
               Trickling Filter Effluent                        4    -     8     mg/L
               Advanced Treatment Effluent              0.1  -     2.0  mg/L

What is normal for one plant is not necessarily normal for another.  Each operator must determine what range reflects normal conditions for his or her plant.  Values outside the established normal range indicate that a condition exists which could yield poor effluent quality.



Quiz 8.1

1.     Why is phosphorus removed from wastewater?

2.     What are the three basic forms of phosphorus found in wastewaters?

3.     What must be done before polyphosphates can be measured?

 

 

Section 8:  TOTAL PHOSPHORUS

DESCRIPTION OF DIGESTION STEP


Since phosphorus exists in several distinct forms in wastewater samples and the approved test method measures only the orthophosphate form, pretreatment methods have been developed to convert the various forms of phosphate-phosphorus to the orthophosphate form.  If the only determination to be made is Total Phosphate-Phosphorus, the sample is digested to convert both the polyphosphate and the organic phosphate to the ortho form at the same time.  If the analyst must test for the various types of phosphate, an acid-hydrolysis must be performed.  Since this is not required for most facilities, it is not included in this text.

Section 9:  EQUIPMENT AND REAGENTS

EQUIPMENT

1.     Hot plate

2.     Tongs or gloves

3.     Scoop (0.4 g capacity)

4.     125 mL Erlenmeyer flasks (acid washed)

5.     50 mL graduated cylinders (acid washed)

REAGENTS

1.     Phenolphthalein indicator

2.     Sulfuric acid solution

3.     Ammonium persulfate, crystal

4.     Sodium hydroxide, 1 N

See Appendix B for the procedures for preparation of the reagents used in this method.

 

Section 10:  LABORATORY PROCEDURE


1.     Measure 50 mL or an appropriate amount of sample diluted to 50 mL with distilled water.

2.     Add 1 drop phenolphthalein indicator.  If a red color develops, add sulfuric acid solution until color just disappears.

3.     Add 1 mL of sulfuric acid solution and 0.4 g of ammonium persulfate. 

4.     Boil gently for 30 to 40 minutes or until the total volume is 10 mL.

5.     Cool, add 1 drop of phenolphthalein and neutralize to a faint pink color with 1 N sodium hydroxide.

6.     Make up to 50 mL with distilled water.  The digested sample is then tested for total phosphate as outlined in Section 12.

NOTE:  If precipitate forms, do not filter.

Section 11:  TOTAL PHOSPHORUS

ASCORBIC ACID PROCEDURE

An accepted method for determining phosphate-phosphorus is the Ascorbic Acid Procedure.  The procedure is suitable for concentrations of 0.01 to 6 mg/L PO4-P.

Ammonium molybdate and potassium antimonyl tartrate react in acidic solution with orthophosphate to form a heteropolyic acid (phosphomopydbic acid) which can be reduced by ascorbic acid to form an intense blue color.

Section 12:  EQUIPMENT AND REAGENTS

EQUIPMENT

1.     Spectrophotometer with an infrared phototube for use at 880 nm with a light path of at least 2.5 cm (1 in)

2.     Acid washed glassware

3.     An analytical balance capable of weighing to 0.1 mg accuracy

REAGENTS

1.     Sulfuric acid solution 5 N

2.     Potassium antimonyl tartrate solution

3.     Ammonium molybdate Solution

4.     Ascorbic acid solution

5.     Combined reagent

6.     Standard phosphate solution

See Appendix B for the procedures for preparation of the reagents used in this method.

Section 13:  LABORATORY PROCEDURE


1.     Pipette 50.0 mL or an appropriate amount diluted to 50 mL of digested sample into an acid cleaned, dry 125 mL Erlenmeyer flask.

NOTE:  If only orthophosphate is to be determined, an undigested sample is used.

2.     Add 1 drop of phenolphthalein indicator.  If a red color develops, add 5 N sulfuric acid until the color disappears.

3.     Add 8.0 mL of combined reagent and mix thoroughly.

4.     Allow at least 10 minutes (but not more than 30 minutes) for color development.

5.     Measure absorbance at 880 nm using a reagent blank to zero the spectrophotometer.

The reagent blank is made using 50 mL of distilled water carried through the digestion step and ascorbic acid procedure.

In highly colored or turbid samples, prepare a sample blank by adding all the reagents to the sample except the ascorbic acid and potassium antimonyl tartrate.  Subtract the absorbance of this blank from the absorbance of the sample.

6.     Check the sample’s absorbance against the calibration curve and determine the concentration.  Correct for dilution.

Section 14:  CALIBRATION CURVE

Since the phosphate concentration is measured as a function of absorbance, a standard curve of absorbance versus known phosphate concentrations must be prepared.  Six standard phosphorus concentrations and a distilled water blank are treated with the same digestion procedures as the samples.  These 6 values are used to plot absorbance versus phosphate concentration to give a straight line passing through the origin.

Prepare 6 dilutions of the 5.0 mg/L Phosphate standard to result in the following final concentrations:

                            final conc.                  volume of std.             final vol. in flask

                              0.1 mg/L                         1.0 mL                         50.0 mL
                              0.2 mg/L                         2.0 mL                         50.0 mL
                              0.4 mg/L                         4.0 mL                         50.0 mL
                              0.6 mg/L                         6.0 mL                         50.0 mL
                              0.8 mg/L                         8.0 mL                         50.0 mL
                              1.0 mg/L                        10.0 mL                        50.0 mL

Perform the Total Phosphorus procedure including the digestion step and color development.  Record the absorbances and plot the curve.  At least one standard phosphate concentration must be included with each batch of samples as a check on the calibration curve.  Prepare a new curve every six months.
See Appendix D for more information on the preparation of calibration curves.

Section 15:  CALCULATIONS

 

Determine the final concentration of the sample using the following equation:

mg/L P  =  mg/L from the curve x 50 mL

divided by initial volume used (mL)

For example, if 5 mL of Influent gives an absorbance of 0.35 which represents 0.558 mg/L on the calibration curve, the corrected concentration of phosphate-phosphorus is:

mg/L P  =  0.558 x 50 mL divided by 5 mL = 5.58 mg/L

Section 16:  INTERFERENCES


Arsenates react with the molybdate reagent to produce a blue color similar to that formed in the phosphate determination.  Concentrations as low as 0.1 mg As/L interfere with the analysis.  Hexavalent chromium and nitrite interfere to give low results.

Chapter 17:  QA/QC


A Quality Assurance/Quality Control program is required by the NPDES permit.  Quality Assurance (QA) is a set of operating principles that must be followed during sample collection and analysis.  Lab bench sheets must be maintained that document when the sample was collected, how it was preserved, and what results were obtained.

Quality Control (QC) includes any testing which is done to prove that the results are reliable.  One of every ten samples analyzed should be a QC check.  This may include duplicate samples, spike samples, reagent blank analyses and known QC samples obtained from outside sources.

Duplicate sample analysis involves analyzing the same sample twice and comparing the results.  The closer the results, the more accurate the analysis.  Results should not differ by more than 10%.  Spike sample analysis involves adding known amounts of analyte to a sample and calculating the percent recovery.  These are discussed further in Chapter 10.

In Phosphorus analysis, a distilled water blank must be run with every batch of samples tested.  It is used both to zero the spectrophotometer and to show that the glassware and reagents are not contaminated with phosphates.

At least one check standard must also be run with each batch.  If a variety of samples are being tested with results expected to cover a wide range of concentrations, test two or more standards.  The absorbencies of the standards must agree closely with the true value or a new curve must be drawn.

Spike samples are run by adding the 5.0 mg/L Phosphate standard to a sample.  To simplify the calculation, use the same amount of standard in the spike sample as you are using in your check standard.

For example, an Effluent sample is spiked with 2 mL of the 5.0 mg/L Phosphate standard yielding the following results:

Effluent sample result:  0.116 mg/L
Spiked Effluent sample result:  0.310 mg/L
Check standard (2 mL) result:  0.197 mg/L

Expected spike result = sample result + standard result


= 0.116 + 0.197
= 0.313

Percent recovery = spike result divided by expected result x 100%

= 0.310 divided by 0.313 x 100%
= 99%

QC samples with known concentrations of Phosphate can be purchased from chemical supply companies.  The results from duplicate, spike and outside QC samples should be recorded in a QC notebook.

Sample bench sheets are included in Appendix C.


Quiz 8.2

1.     What four reagents are used to make the combined reagent for the phosphorus test?  What is the storage time of this reagent?

2.     What sample volume would you start with if the estimated phosphorus content of the sample was 1 mg/L?  4 mg/L?  10 mg/L?





Answers to Quizzes

Quiz 8.1


1.     Why is phosphorus removed from wastewater?

Phosphorus is removed from wastewater because it provides a nutrient or food source for algae.  Dead algae can cause serious oxygen depletion problems in receiving streams which in turn can kill fish and other aquatic life.  Also, algae can cause taste and odor problems in drinking water supplies.

2.     What are the three basic forms of phosphorus found in wastewaters?

        The three basic forms of phosphorus found in wastewaters are Orthophosphates Polyphosphates, and Organic phosphates.

3.     What must be done before polyphosphates can be measured?

The samples must be acid hydrolyzed (or digested) to convert the polyphosphates to the orthophosphate form for measurement.

Quiz 8.2

1.     What four reagents are used to make the combined reagent for the phosphorus test?  What is the storage time of this reagent?


        The four reagents that are used to make the combined reagent for the phosphorus test are:

        5 N Sulfuric acid solution;
        Potassium antimonyl tartrate solution;
        Ammonium molybdate Solution; and,
        Ascorbic acid solution.
        The storage time is 4 hours.

2.     What sample volume would you start with if the estimated phosphorus content of the sample was 1 mg/L? 4 mg/L? 10 mg/L?


        20 to 30 mL for 1 mg/L
        5 to 10 mL for 4 mg/L
1 to  4 mL for 10 mg/L
 

APPENDIX A
 
References


Standard Methods for the Examination of Water and Wastewater, 18thth Edition, AWWA, APHA, WPCF; Water Pollution Control Federation, Washington, DC, 1992.

Methods for Chemical Analysis of Water and Wastes, U.S. EPA - 600/4-79-020, March 1979.

A Field Study Program, Operation of Wastewater Treatment Plants, Kerri, Kenneth et al, University of California, Sacramento.

NOTES:


APPENDIX B

Phosphorus Reagents

Digestion reagents:

1.     Sulfuric acid digestion reagent


Carefully add 300 mL concentrated Sulfuric acid to approximately 600 mL distilled water and dilute to 1 L with distilled water.

2.     Sodium hydroxide  1 N


Dissolve 40 g of Sodium hydroxide pellets in distilled water.  Cool and dilute to 1 L.

Ascorbic acid procedure:

1.     Sulfuric acid solution  5 N

Dilute 70 mL of concentrated Sulfuric acid to 500 mL with distilled water  (add acid to water, CAREFULLY).

2.     Potassium antimonyl tartrate solution

Dissolve 1.3715 gm of Potassium antimonyl tartrate in 400 mL of distilled water in a 500 mL volumetric flask and dilute to 500 mL.  Store in a glass stoppered bottle.

3.     Ammonium molybdate solution

Dissolve 20 g of Ammonium molybdate in 500 mL of distilled water.  Store in a glass stoppered bottle.
 
4.     Ascorbic acid solution  0.1 M

Dissolve 1.76 g of Ascorbic acid in 100 mL of distilled water.  Store at 4°C and discard after 1 week.

5.     Combined reagent

To prepare 100 mL of combined reagent, combine the following reagents and mix thoroughly after each addition.

50 mL 5 N Sulfuric acid
5 mL Potassium antimonyl tartrate solution
15 mL Ammonium molybdate solution
30 mL Ascorbic acid solution

Reagents must be added in the order listed.  All reagents must be at room temperature before mixing.  If turbidity occurs, shake the reagent and let stand until turbidity clears.

The combined reagent is stable for only 4 hours.

6.     Stock Phosphate Solution

Dissolve 219.5 mg of anhydrous potassium dihydrogen phosphate in distilled water and dilute to 1 L.  This yields a concentration of 50 mg/L.

7.     Standard Phosphate Solution

        Dilute 10 mL of the stock solution to 100 mL with distilled water.  This yields a concentration of 5.0 mg/L.


APPENDIX C

 

Sample Bench Sheet


Phosphate Analysis

Sample collection Date: _______          Time: _______                By: _______
Sample preserved?  yes  no                 Preservative: ____________
Analysis run Date: _______                  Time: _______                By: _______


Sample
Type

Initial
Volume

Final
Volume

Absorbance

Conc.

Corrected
Conc.














































































































APPENDIX D

 

Preparation of a Calibration Curve


In several of the analyses presented in this manual, there is a need for the preparation of a standard calibration curve.  The curve establishes the relationship between absorbance or percent transmittance and the concentration of the desired parameter.  Using this curve, samples can be tested by the measurement of their absorbance or percent transmittance and the corresponding concentration can be read off the graph.

Rules for preparing calibration curves.

1.     Prepare all standards in duplicate or triplicate.  Use the average of absorbance or transmittance for preparing the curve.

2.     Process the standards in the same manner as the samples will be analyzed.  If the samples will
        undergo a digestion, the standards must be digested as well.

3.     A distilled water sample (blank) must be processed in the same manner as the standards.  This blank will be used as the reagent blank for setting the spectrophotometer to read zero absorbance or 100% transmittance.

4.     A minimum of one standard should be processed daily to check the accuracy of the calibration curve and the quality of the reagents.

5.     When new reagents are prepared, the calibration curve should be checked by at least 3 standards.

6.     If the curve appears to differ from a straight line, the curve may be straightened to approximate a straight line by plotting the data on semi-logarithmic graph paper.

7.     Plot data with the absorbance or percent transmittance readings on the vertical axis and concentration on the horizontal axis.

8.     Select graph paper with appropriate graduations.  Do not use graduations which require constant estimation of readings.

No comments:

Post a Comment