Detector Gases
Three gases are required for FID detection with capillary columns :
- A FUEL gas, typically Hydrogen
- An OXIDISER gas, typically Air
- A MAKEUP gas, typically Nitrogen or Helium
The Fuel gas and Oxidiser combine to form the
flame in which the analyte will combust and ionise. The makeup gas
assists the analyte up into the flame and through the detector and
minimizes dead volume effects.
Detector Gas Flow Rates
The ratio’s of detector gas flow rates are
crucial for optimum sensitivity and good peak shape. Most FID detectors
operate at maximum sensitivity with a volumetric flow rate of between
20 and 40 ml/min. This is a combination of carrier and makeup gas
flows. As most capillary columns operate between 0.1 and 10 ml/min,
well below the optimum required for maximum sensitivity, we can set the
makeup gas flow rate accordingly.
Typical FID default flow rates:
Carrier Gas:
Makeup Gas:
Hydrogen:
Air: |
1ml/min.
30ml/min.
30ml/min.
300ml/min |
If the makeup gas flow rate is too low, the
analyte band is not swept effectively through the dead volume of the
detector and peak tailing / broadening will occur. If the makeup gas
flow is too high, then peak shape will be good but sensitivity may be
reduced as the analyte is swept through too quickly for the detector to
respond fully.
|
|
Figure 1: Typical FID detector schematic |
|
|
|
|
FID Sensitivity
The sensitivity of the FID detector depends upon several factors that include:
- Combustion gas flow rates (and their relative ratio sometimes called the Gas Stoichiometry)
- Makeup gas flow rate (see above)
- Flame jet exit diameter
- Carrier Gas Flow Rate
- Relative positions of jet (anode) and collector (cathode)
- Position of the column relative to the jet exit
- Detector temperature (to a lesser extent)
- The optimum stoichiometry of the combustion gases is typically 10:1, air to hydrogen.
|
|
|
|
|
|
|
|
|
(a) The sensitivity of the detector will typically deviate with hydrogen flow as shown in Figure 2.
The FID
detector burns a Hydrogen / Air flame which is hot enough to pyrolise
most organic compounds which pass through it producing cations and
electrons which move between the anode and cathode to form a minute
current. The exact mechanisms of ion production are not well
understood, however the ‘composition’ of the flame, that is the ratio
of air to hydrogen, appears to have a significant effect on ion
production and therefore sensitivity. The response of the detector is
directly proportional to the number of reduced carbon atoms moving
between the two electrodes per unit time.
If hydrogen flow rate is too high this tends to reduce the dynamic
linear range of the detector and one should use the manufacturers
recommended hydrogen flow rate wherever possible as a starting point. |
|
Figure 2: Effect of Hydrogen on
relative sensitivity of FID detector
|
|
|
|
|
(b) Makeup gas flow rate (see above) |
|
|
|
(c) Flame jet exit diameter |
|
The
jet diameter will dictate the rate at which the carrier gar flows into
the flame – i.e. the volumetric flow and linear velocity of the gas
eluting in the flame. With packed GC columns the carrier gas flow
rate (and hence the volumetric flow into the flame) is usually high
enough to produce a reasonable detector response. Standard FID jets have
exit diameters of approximately 0.5–0.7 mm, which is suitable for most
applications.
A smaller jet of about 0.3 mm i.d. is often used with capillary columns
to gain sensitivity (about 1.5x) through increasing linear velocity of
the carrier into the flame. A narrow FID jet is not recommended for
packed-column use because stray column packing support easily can clog
the jet passage. Conversely, narrower jets prevent the tip of a
capillary from accidentally protruding into the flame.
|
|
|
|
|
|
|
(d) Carrier Gas Flow Rate
Capillary GC carrier flow can be
operated in two modes – ‘Constant Pressure’ or ‘Constant Flow’ In
constant pressure mode, the flow of the carrier gas can reduce as the
oven temperature increases. Whilst the
FID detector is not ‘Mass Flow Sensitive’ some baseline rise may be
observed during the temperature gradient.
This phenomenon will be eliminated when
operating in ‘Constant Flow’ mode as the instrument adjusts the
pressure to maintain constant carrier flow during the temperature
gradient. Some modern FID detectors offer the ability to ramp the make
up gas flow so that overall ‘carrier + make up’ flow does not alter
even when operating the carrier gas in constant pressure mode.
(e) Relative positions of jet (anode) and collector (cathode)
The relative
positions of jet (anode) and collector (cathode) are fixed by the
instrument manufacturer to give the optimum sensitivity.
back to top »
(f) Position of the column relative to the jet exit
If the tip
of the capillary column is placed too low in the FID jet, there will be
a large unswept volume and peak broadening will occur.
If the end of the column is touching the jet tip, the polyimide column
coating will ‘bake out’ and may chip, releasing small particles of
polyimide into the flame, causing ‘spikes’ on the baseline.
|
|
Spikes on the GC baseline - potentially caused by
incorrect positioning of the capillary column in the FID detector.
The ideal position for the column end is 1-2mm below the jet orifice. |
(g) Detector temperature (to a lesser extent)
The
response (sensitivity) of the detector is not closely linked with
detector temperature providing two basic conditions are met. First,
the detector must be at a minimum of 150oC for stable operation and to
avoid water condensation in the detector housing; and second that the
detector temperature is between 20 and 50oC above the maximum column
oven temperature for that particular application. Any higher than
this and stationary phase may bleed from the portion of the column
inside the detector causing baseline disturbances, noise and
potentially poor peak shape due to analyte adsorption on exposed silica
surfaces.
|
|
|
|
|
|
|
|
|
The FID
detector is highly robust and reliable with a wide linear range.
However, as with most GC detectors, there is a reliance on high purity
bases and routine maintenance to ensure optimal performance. We have
highlighted here some of the key maintenance and troubleshooting
operations, which will be further highlighted during the next Essential
Guide Webcast on April 26th. |
(a) Gas supplies:
All detector gases must meet minimum purity
and flow-control specifications. Table I summarizes carrier,
combustion, and makeup gas purity, purification, and regulation
requirements for some common detector systems.
In addition to
procuring sufficiently pure gases, chromatographers must ensure that
appropriate gas pressure regulators and purifiers are in place. An
inexpensive regulator can add impurities to the gas passing through and
fail to regulate the gas pressure accurately. Without proper
purification, minute leaks between a tank and an instrument can
compromise gas purity and eliminate the advantage of more-expensive,
high-purity gases.
|
No comments:
Post a Comment