Standard
operating procedure of calibration of Oven using calibrated thermometer.
1.0
OBJECTIVE
To describe the
procedure for operation and calibration of precision oven.
2.0 SCOPE
This SOP is
applicable for operation and calibration of precision oven.
3.0
RESPONSIBILITY
Officer/
Executive - Quality Control
4.0 ACCOUNTABILITY
Manager -
Quality Control
5.1
Operation
5.1.1 Ensure that temperature indicator is
within due date of calibration.
5.1.2 Switch on the precision oven.
5.1.3 Adjust the required
temperature by adjusting temperature knob (soft key).
5.1.4 After attaining required
temperature keep the dish and lid separately inside the oven. Close tightly the
oven door.
5.1.5 Allow to hold the
specified temperature.
5.1.6 Note the time of
attaining specified
temperature.
5.1.7 After completion of specified
time switched off the precision oven.
5.1.8 Allow to come down the
temperature below 40°C.
5.1.9 After attaining to 40°C,
open the door and take out the dish & replace the lid.
5.1.10 Reclose the door.
5.1.11 Take the print out of
the duration used for LOD test.
5.1.11 Fill the usage log
book.
5.2 Calibration
5.2.1 Calibrate the chamber
temperature by calibrated thermometer.
5.2.2 Set the chamber
temperature alternately 50° C, 105° C and 150° C, Standardize by calibrated
thermometer. Record the observations.
5.2.3 Calibrate the
temperature sensor by the external agency on annual basis.
5.2.4 Calibration frequency of
chamber temperature: Quarterly.
5.3 Precautions
5.3.1 Never keep switch ON
position, when not in use.
5.3.2 Clean equipment after
use
5.4 Maintenance
5.4.1 If the instrument does
not produce required results or its response is poor then it should be labeled
“UNDER MAINTENANCE” and should be repaired or serviced.
6.0
ABBREVIATIONS:
6.1 SOP - Standard Operating
Procedure
6.2 No. - Number
6.3 QA - Quality Assurance
6.4 NA - Not Applicable
6.5 CCF - Change control
format
6.6
QC - Quality Control
ANNEXURE-I
Format No.
PRECISION OVEN USAGE LOG
Format No.
PRECISION OVEN USAGE LOG
Date
|
Product/
Material Name
|
Batch No./ Lot No.
|
Time
|
Analyzed
by
|
Checked by
|
Remarks
|
|
Started
at
|
Completed
on
|
||||||
ANNEXURE-II
Format No.
CALIBRATION REPORT OF PRECISION OVEN
Asset ID No.
|
:
|
Ref. SOP No.
|
:
|
||
Make
|
:
|
Model No.
|
:
|
1. Cleanliness of the Instrument:
Observation: Clean / Not Clean
Corrective action in the case of non-compliance:
2. Calibration :
Observations:
Calibrated thermometer ID No.:__________________
Observation: Clean / Not Clean
Corrective action in the case of non-compliance:
2. Calibration :
Observations:
Calibrated thermometer ID No.:__________________
S. No.
|
Temperature
|
Limit
|
Remarks
|
|
Set
|
Observed
|
|||
50°C
|
+ 20 C
|
|||
105°C
|
+ 20 C
|
|||
150°C
|
+ 20 C
|
|||
Remarks: The
instrument is calibrated as per the SOP and shows satisfactory / does not satisfactory results as per
calibration procedure.
Calibrated
on : _______________
Next Calibration Due on :________________
Next Calibration Due on :________________
Done by: Checked
by:
Date:
Date: